HomeEmail

ZEON
Critical Care Ventilator

ZEON Critical Care Ventilator
  • Suitable for critical care applications
  • Microprocessor controlled, electrically driven ventilator
  • Large LCD display
  • In-built battery backup
  • User friendly controls
  • Low operating cost
  • Application: ICU and ICCU

Technical Specifications

Specifications

Input Supply 185 - 264 V, 50/60 Hz, AC
Power Consumption 40 Watts
VA Requirement 70 VA
Internal Battery 2 X 6 V, 4.5 AH
Battery Back-up Duration 60 minutes
Dimensions 480 mm (W) x 350 mm (D) x
285 mm (H)
Weight 20 Kg
Operating Temperature 10 - 40oC

Operational Parameters

Modes CMV-VC, ASSIST-VC, SIMV-VC, PSV-ASSIST, SIMV-PSV,PSV APB, CMV-PCV, ASSIST-PCV, SIMV-PCV
Breaths per Minute (1 to 60 - SIMV) 6 to 60
Tidal Volume 30 - 1100 cc
Inspiratory Pressure 10 to 60 cm H2O
Flow Pattern Constant
I:E Ratio 1:4 to 4:1
Inspiratory Time 0.04 to 4 sec (20% to 80% in CMV)
Pressure Trigger -15 to -0.5 cm H2O
Flow Trigger 1.8 to 30 lpm
Peep Level 0 to +20 cm H2O
PCV Limit 10 to 60 cm H2O
PSV LImit 10 to 60 cm H2O
Apnea Backup Time 5 to 60 secs
Other Features Inspiratory Pause Limit, Sigh Breath, Manual Breath

Alarms/Indications

  • » Excess Breaths
  • » Low Expired Volume
  • » Low Inspired Volume
  • » FiO2 Sensor Open
  • » FiO2 Sensor Defective
  • » FiO2 High
  • » FiO2 Low
  • » High Inspiratory Pressure
  • » Low Inspiratory Pressure
  • » Patient Disconnection
  • » Ventilator Malfunction
  • » Low Inlet Volume
  • » Check Exp Flow Sensor
  • » Apnea
  • » Power Failure
  • » Peep Low
  • » Peep High
  • » Battery Operation
  • » Battery Low/ Battery Discharged
  • » Ventilator Failure

Monitored Parameters

  • » FiO2 Monitoring
  • » Spontaneous Breaths
  • » Inspired Tidal Volume
  • » Expired Tidal Volume
  • » Peak Inspiratory Pressure
  • » Plateau Pressure
  • » PEEP Level
  • » Trigger Level
  • » Delivered BPM

Connections

  • » 22 mm (Male) outlet port for tubing connection to patient.
  • » 22 mm (Male) expiratory port for tubing connection from patient.
  • » 8 mm low pressure connection for oxygen enrichment.
  • » Pressure monitoring port.
  • » Reservoir Bag Port